Consent Preferences
WarningFlammable

9400

Transmisor de nivel de radar de ondas guiadas (GWR)

Application
The 9400 series guided wave radar level transmitter is a breakthrough in cost effective very high accuracy, very high resolution liquid level sensing for heavy duty applications. Rochester Sensors’ GWR level sensor is the culmination of years of research into micropower impulse radar, time domain reflectometry and equivalent time sampling. Rochester Sensors Model 9400 liquid level sensor is used to monitor liquids such as fuels, chemicals, water, coolant, lubricants and other materials. The device is capable of maintaining very accurate level readings regardless of liquid. The 9400 can optionally be configured to accurately measure both fuel level and the level of dissimilar materials, such as water/fuel interface. The device emits a series of electromagnetic pulses down the sensor probe. Liquid/air interface, and dissimilar liquid interfaces provide a change in the characteristic impedance of the sensor probe, and create a reflection of the pulse. The time-of-flight of the reflected pulses is measured, and the liquid and interface levels are calculated.

  • Los vehículos comerciales y autobuses
  • Maquinaria de construcción y minería
  • Maquinaria agrícola
  • Generadores de energía móviles y estacionarios, bombas y compresores

Como se muestra en la siguiente tabla, el modelo 9400 tiene muchas ventajas que otras tecnologías no tienen y tiene un costo similar a las soluciones ultrasónicas o capacitivas, pero mucho más barato que un típico producto de radar de ondas guiadas para aplicaciones de control de procesos.

Documents
9400 Introduction
Data Sheet


Especificaciones generales

Active Sensing Range
15mm to 1000mm from flange
Housing Size
45mm (Height) x 67mm (diameter)
Connector
Configurable depending on output requirement
Operating and Storage Temperature Range
-40ºC to +105ºC
Water and Dust Sealing
IP69K
Maximum Tank Pressure
15 PSI
Shock
ISO 16750-3
Vibration
BS EN 60068-2-64: 1993 IEC 60068-2
Up to 500mm
Drop
ISO 16750-3
Surface Resistance
ISO 16750-3
Chemical Resistance
ISO 16750-5
Corrosion Resistance
VDA 621-415, DIN EN ISO 16270-2, DIN EN ISO 9227
Corrosive Gas Resistance
DIN 50018

Especificaciones eléctricas

Supply Voltage
9-36 Vdc or 5Vdc regulated
Supply Current
35mA average
Superimposed AC over DC
ISO 16750-2
Supply Voltage Drop
ISO 16750-2 Section 4.6
EMC/EMI
UN ECE R10 revision 5

Especificaciones de salida

Output Protocol
Analog – Voltage, PWM; Digital – LIN, CAN; Optional
programmable alarm output
Accuracy
±2.0% of span max (including nonlinearity,
hysteresis, temperature effects).
Resolution
1mm

Construcción

Housing
Die Cast Low Copper Aluminum
Sensor Tube
Extruded Low Copper Aluminum
Wetted Materials
Low Copper Aluminum, Buna Nitrile (or Viton), FR-4
Mounting Configuration
SAE 5 Bolt
Weight
Configurable including 5 and 6 bolt SAE

* Los materiales y especificaciones están sujetos a cambios sin previo aviso. Las clasificaciones de presión están sujetas a cambios debido a la temperatura y otras consideraciones medioambientales.

Consulta del producto-9400

    Revisado 04/8/2019